Immobilised activated sludge based biosensor for biochemical oxygen demand measurement |
| |
Authors: | Liu J Björnsson L Mattiasson B |
| |
Affiliation: | Department of Biotechnology, Centre for Chemistry and Chemical Engineering, Lund University, P.O. Box 124, S-221 00 Lund, Sweden |
| |
Abstract: | A biochemical oxygen demand (BOD) sensor, based on an immobilised mixed culture of microorganisms in combination with a dissolved oxygen electrode, has been developed for the purpose of on-line monitoring of the biological treatment process for waste and wastewater. The sensor was designed for easy replacement of the biomembrane, thereby making it suitable for short-term use. The drawbacks of activated sludge based sensor, such as short sensor lifetime, were thereby circumvented. The sensor BOD measurements were carried out in the kinetic mode using a flow injection system, resulting in 25 s for one measurement followed by 4–8 min recovery time. Based on the results of normalised sensor responses, the OECD synthetic wastewater was considered to be a more suitable calibration solution in comparison with the GGA solution. Good agreement was achieved between the results of the sensor BOD measurement and those obtained from BOD5 analysis of a wastewater sample from a food-processing factory. Reproducibility of responses using one sensor was below ±5.6% standard deviation. Reproducibility of responses using different sensors was within acceptable bias limits, viz. ±15% standard deviation. |
| |
Keywords: | BOD sensor Biochemical oxygen demand Activated sludge Microbial consortium Kinetic mode Dynamic mode Wastewater |