首页 | 本学科首页   官方微博 | 高级检索  
     


An unusual cause of recurrent syncope in a patient with implantable cardioverter defibrillator
Authors:Armando Gardini  Francesca Bellandi  Francesco Fracassi
Affiliation:Division of Cardiology, Istituto Clinico S. Rocco, Via dei Sabbioni 24, Ome, Brescia, Italy
Abstract:Electromagnetic interferences (EMI) deriving from electrical devices may affect implantable cardioverter defibrillators (ICD). Improved algorithms have been developed in order to minimize adverse effects. However, caution should be still recommended in ICD recipients when handling electrical devices. Here we describe the case of an ICD patient with recurrent syncopal episodes due to inhibition of pacing by oversensing of electrical noise from a not properly grounded washing machine.
Keywords:Implantable cardioverter defibrillator   Electromagnetic interference   Syncope