首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of plasma histamine by stable isotope dilution gas chromatography-mass spectrometry: methodology and normal values
Authors:J J Keyzer  B G Wolthers  F A Muskiet  H Breukelman  H F Kauffman  K de Vries
Affiliation:1. Central Laboratory for Clinical Chemistry, University Hospital, Groningen, The Netherlands;2. the Department of Allergology, University Hospital, Groningen, The Netherlands
Abstract:A new method for the determination of histamine by stable isotope dilution mass fragmentography is described. The method is specific, sensitive, and accurate, resulting in a within-day coefficient of variation of 4.1% and a day-to-day variation of 7.9%. It was shown that the first blood sample after a venipuncture can contain an artificially elevated plasma histamine concentration. Platelets contain about 7 pmol histamine/10(9) cells. Serum histamine was elevated about four times in comparison with plasma histamine. This phenomenon was mainly ascribed to degranulation of basophilic leukocytes by complement activation during blood clotting. Normal values for plasma histamine were (n = 25) 2.07 +/- 0.75 nmol/liter (mean +/- 1 SD), which is one of the lowest values reported up to now.
Keywords:histamine  gas chromatography-mass spectrometry  plasma histamine  serum histamine  normal values  mastocytosis
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号