首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Accuracies in Po-210 determination for lead-210 dating
Authors:F El-Daoushy  K Olsson  R Garcia-Tenorio
Institution:(1) Department of Physics, Box 530, S-751 21 Uppsala;(2) Present address: Department of Quaternary Geology, Box 555, S-751 22 Uppsala, Sweden;(3) Present address: Facultad de Fisica, Apartado 1065, E-410 80 Sevilla, Spain
Abstract:Various laboratory techniques have been utilized worldwide for measuring lead-210 in sub-recent deposits through its grand-daughter product polonium-210. Isotope dilution alpha spectrometry proved a suitable tool for absolute determination of lead-210 for the dating of aquatic deposits. Moreover, isotope dilution alpha spectrometry along with speciation experiments can be used to resolve depositional anomalies arising from supported lead-210/Ra-226 disequilibrium levels and unsupported lead-210 mobile fractions. Isotope dilution alpha spectrometry of sub-recent sediment and peat deposits has been critically evaluated for more than ten years. Our results show that type, size and composition of deposits analyzed as well as radiochemical procedures used, together with alpha counting techniques, are important factors influencing lead-210 determinations and tailing corrections using its granddaughter product polonium-210. Optimization of these parameters is of prime importance to achieve economic and accurate analyses, especially at low lead-210 concentrations and small sample sizes.
Keywords:aquatic deposits  210Pb/210Po  speciation  dating  alpha spectrometry  accuracies in alpha-counting
本文献已被 SpringerLink 等数据库收录!