首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Degradation of dansyl polyamines on high-performance thin-layer chromatographic plates
Institution:1. Facultad de Ciencias Biológicas, Universidad de Concepción, Concepción, Chile;2. Facultad de Farmacia, Universidad de Concepción, Concepción, Chile;3. Facultad de Ciencias Veterinarias, Universidad de Concepción, Concepción, Chile;4. Faculty of Science, Engineering and Computing, Kingston University, London, United Kingdom.;5. Laboratorio de Inmunobiotecnología, Centro de Referencia para Lactobacilos (CERELA-CONICET), Tucumán, Argentina
Abstract:Using high-performance thin-layer chromatography with in situ quantitation to measure dansylated polyamines in the range of 1–20 pmol, we found that dansylated polyamines apparently react with the silica gel of the plates. The fluorescence of the dansyl polyamines diminished with increase in the time interval between application of a sample to the plate and start of the chromatographic separation. Conversely, the fluorescence at the site of application increased with the length of the time interval, indicating the formation of polar reaction products. If this reaction is not accounted for, considerable errors in quantitation of dansyl polyamines may occur.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号