An in vivo method for measuring turbulence in mechanical prosthesis leakage jets
Authors:Travis Brandon R  Christensen Thomas D  Smerup Morten  Olsen Morten S  Hasenkam J Michael  Nygaard Hans
Affiliation:Department of Cardiothoracic and Vascular Surgery, Arthus Universitetshospital, Skejby Sygehus, Brendstrupgaardsvej, 8200 Arhus N, Denmark. brandontravis2001@yahoo.com
Abstract:This work introduces a method for the in vivo measurement and analysis of turbulence within the leakage of a mechanical heart valve. Several analysis techniques were applied to ultrasound measurements acquired within the atrium of a pig, and error associated with these techniques was analyzed. The technique chosen applies cyclic averaging to mean and maximum velocity measurements within small, normalized phase windows to calculate Reynolds normal stresses in the direction of the ultrasound beam. Maximum shear stresses are estimated from these normal stresses using an analytical technique. The stresses observed were smaller than those reported from previous in vitro simulations.