Criteria for set point estimation in the volume clamp method of blood pressure measurement.
Authors:J Penáz
Institution:Department of Physiology, Medical Faculty, Masaryk University, Brno, Czechoslovakia.
Abstract:When blood pressure is measured in the finger using the volume clamp method the value at which the vascular volume is clamped is of crucial importance. Since the discovery of the method, several criteria of finding a correct set point have been elaborated: 1. The volume oscillations reach their maximum amplitude at cuff pressure equalling mean blood pressure. 2. The form of the diastolic portion of volume pulsations changes if the cuff pressure moves around the mean blood pressure. 3. The set point can be positioned at one third of the arterial volume. 4. The dynamic vascular compliance (DVC) may be continuously measured as the instantaneous amplitude of vascular volume oscillations is elicited by a relatively small and rapid vibration of the cuff pressure. The shape of the DVC pulse characteristically depends on the transmural pressure (TP): at negative TP (cuff pressure exceeding the blood pressure) it shows a distinct positive systolic peak, at positive TP the polarity of the DVC pulse is reversed. In contrast to the first three ways to find the set point, the last one may operate even in closed-loop performance, i.e. during the blood pressure measurement.