A contact lens with embedded sensor for monitoring tear glucose level
Authors:Yao Huanfen  Shum Angela J  Cowan Melissa  Lähdesmäki Ilkka  Parviz Babak A
Institution:Department of Electrical Engineering, University of Washington, Seattle, WA 98195, USA.
Abstract:We report the design, construction, and testing of a contact lens with an integrated amperometric glucose sensor, proposing the possibility of in situ human health monitoring simply by wearing a contact lens. The glucose sensor was constructed by creating microstructures on a polymer substrate, which was subsequently shaped into a contact lens. Titania sol-gel film was applied to immobilize glucose oxidase, and Nafion? was used to decrease several potential interferences (ascorbic acid, lactate, and urea) present in the tear film. The sensor exhibits a fast response (20s), a high sensitivity (240 μA cm(-2) mM(-1)) and a good reproducibility after testing a number of sensors. It shows good linearity for the typical range of glucose concentrations in the tear film (0.1-0.6 mM), and acceptable accuracy in the presence of interfering agents. The sensor can attain a minimum detection of less than 0.01 mM glucose.