首页 | 本学科首页   官方微博 | 高级检索  
     


A new method for evaluation of cavitation near mechanical heart valves
Authors:Johansen Peter  Manning Keefe B  Tarbell John M  Fontaine Arnold A  Deutsch Steven  Nygaard Hans
Affiliation:Bioengineering Department, Penn State University, University Park, PA 16802, USA. peter.johansen@iekf.au.dk
Abstract:Evaluation of cavitation in vivo is often based on recordings of high-pass filtered random high-frequency pressure fluctuations. We hypothesized that cavitation signal components are more appropriately assessed by a new method for extraction of random signal components of the pressure signals. We investigated three different valve types and found a high correlation between the two methods (r2: 0.8806-0.9887). The new method showed that the cavitation signal could be extracted without a priori knowledge needed for setting the high-pass filter cut off frequency, nor did it introduce bandwidth limitation of the cavitation signal.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏