首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Simple high-performance liquid chromatographic method for determination of ketoconazole in human plasma
Authors:Kah Hay Yuen  Kok Khiang Peh
Institution:School of Pharmaceutical Sciences, University of Science Malaysia, 11800 Penang, Malaysia
Abstract:A simple high-performance liquid chromatographic method using fluorescence detection was developed for the determination of ketoconazole in human plasma. The method entailed direct injection of the plasma sample after deproteinization using acetonitrile. The mobile phase comprised 0.05 M disodium hydrogen orthophosphate and acetonitrile (50:50, v/v) adjusted to pH 6. Analysis was run at a flow-rate of 1.5 ml/min with the detector operating at an excitation wavelength of 260 nm and an emission wavelength of 375 nm. The method is specific and sensitive with a quantification limit of approximately 60 ng/ml and a detection limit of 40 ng/ml at a signal-to-noise ratio of 3:1. Mean absolute recovery value was about 105%, while the within-day and between-day coefficient of variation and percent error values of the assay method were all less than 14%. The calibration curve was linear over a concentration range of 62.5–8000 ng/ml.
Keywords:Ketoconazole
本文献已被 ScienceDirect 等数据库收录!