Affiliation: | 1. School of Biomedical Engineering (Suzhou), Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, China Jiangsu Key Laboratory of Medical Optics, Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Science, Suzhou, China;2. Jiangsu Key Laboratory of Medical Optics, Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Science, Suzhou, China;3. School of Biomedical Engineering (Suzhou), Division of Life Sciences and Medicine, University of Science and Technology of China, Hefei, China |
Abstract: | In this work, we developed a motion estimation and correction method which real-time obtained the direction and displacement of repetitive micro bulk motion (such as cardiac and respiratory motion) on an SS-OCT system without additional tracking hardware, and reduced the motion noise in optical coherence tomography angiography (OCTA). In the approach, the direction of repetitive micro bulk motion was considered fixed, and proportional relationships between the motion components in three directions were determined; Then we performed one-dimension cross-correlation to obtain depth displacement which was further used to obtain other two motion components, and greatly reduced the computation; The processing speed on a graphic processing unit was 478 pairs of B-Scans per second, and the measurement range was larger than the range of the angiogram-based methods. Lastly, corrupt angiograms were recovered by adaptive scan protocol, and reduced acquisition time in comparison with the previous work. |