首页 | 本学科首页   官方微博 | 高级检索  
     


Piezoelectric immunosensor for the direct and rapid detection ofFrancisella tularensis
Authors:M. Pohanka  P. Skládal
Affiliation:1.Center of Biological Defense,Těchonín,Czechia;2.Department of Biochemistry, Faculty of Science,Masaryk University,Brno,Czechia
Abstract:A novel immunosensing device based on a piezoelectric sensor for direct detection of the biological warfare agent Francisella tularensis was developed. This sensor includes mouse polyclonal antibody immobilized in a layer of protein A covalently linked to the gold electrode of the sensor. The immunosensor is able to detect F. tularensis with the limit of detection 10(5) CFU/mL with a typical measuring cycle > 5 min. The sensor was successfully evaluated for rapid detection of F. tularensis spikes in drinking water and milk; no deterioration of sensitivity in comparison with buffer solutions was observed. The proposed concept of a rapid measurement of microbial agents seems to be promising for evaluation of samples after short pre-cultivation enrichment.
Keywords:
本文献已被 PubMed SpringerLink 等数据库收录!