Abstract:The recent implication of genomic copy number variations (CNVs) in multiple human genetic disorders has led to increased interest in CNV discovery technologies. There is a growing consensus that, in addition to the method used for detection, at least one additional technology should be employed for validation. Real-time quantitative polymerase chain reaction (qPCR) analysis, incorporating a normal (2N) copy number standard, is commonly used as a means of validating CNVs. Whereas it has previously been reported that formalin-fixed paraffin-embedded (FFPE) DNA samples can yield spurious CNV calls in real-time qPCR assays, here we report that sample degradation under standard laboratory storage conditions generates a significant increase in false-positive CNV results. Results suggest the possibility of biased degradation among genomic regions and emphasize the need to assess sample integrity immediately prior to real-time qPCR experiments.