首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Microrecirculation reactor system for characterization of immobilized enzymes
Authors:James B Taylor  Harold E Swaisgood
Abstract:A differential microrecirculation reactor was developed for kinetic analysis of both soluble and immobilized enzymes. The reactor system was easliy fabricated with in the laboratory from readily available materials. The disposable, small reactors allowed for in situ weight determination of the enzyme beads. Routinely, only a 1 ml liquid volume of substrate was used for each kinetic assay. The reactor system was also used for determination of partition coefficients. Both uses of the reactor system required only 5–10 min for completion of a given determination.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏