Polyphenol biosensor based on laccase immobilized onto silver nanoparticles/multiwalled carbon nanotube/polyaniline gold electrode |
| |
Authors: | Rawal Rachna Chawla Sheetal Pundir C S |
| |
Affiliation: | Department of Biochemistry, M.D. University, Rohtak 124 001, India |
| |
Abstract: | Laccase purified from Ganoderma sp. was immobilized covalently onto electrochemically deposited silver nanoparticles (AgNPs)/carboxylated multiwalled carbon nanotubes (cMWCNT)/polyaniline (PANI) layer on the surface of gold (Au) electrode. A polyphenol biosensor was fabricated using this enzyme electrode (laccase/AgNPs/cMWCNT/PANI/Au electrode) as the working electrode, Ag/AgCl as the reference electrode, and platinum (Pt) wire as the auxiliary electrode connected through a potentiostat. The biosensor showed optimal response at pH 5.5 (0.1 M acetate buffer) and 35 °C when operated at a scan rate of 50 mV s−1. Linear range, response time, and detection limit were 0.1–500 μM, 6 s, and 0.1 μM, respectively. The sensor was employed for the determination of total phenolic content in tea, alcoholic beverages, and pharmaceutical formulations. The enzyme electrode was used 200 times over a period of 4 months when stored at 4 °C. The biosensor has an advantage over earlier enzyme sensors in that it has no leakage of enzyme during reuse and is unaffected by the external environment due to the protective PANI microenvironment. |
| |
Keywords: | Polyphenol Laccase Amperometric biosensor Ganoderma sp. AgNPs cMWCNT |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|