首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Minimizing matrix effects in the development of a method for the determination of salmeterol in human plasma by LC/MS/MS at low pg/mL concentration levels
Authors:Cápka Vladimír  Carter Spencer J
Institution:Tandem Labs, 1121 East 3900 South, Salt Lake City, UT 84124, USA. vladimir@tandemlabs.com
Abstract:Salmeterol is an inhaled bronchodilator drug used for treatment of asthma. Its concentrations in plasma are very low or undetectable by previously developed methods. The present paper describes a method for analysis of salmeterol in human plasma with 2.5 pg/mL lower limit of quantitation. Despite the basic character of the drug the method uses mixed mode anion-exchange solid phase extraction for sample preparation combined with a column switching approach to minimize matrix effects. Samples are separated and detected by LC/MS/MS. The method is easy to use, only requires 0.5 mL of plasma and was validated for use in bioanalytical applications. The method does not suffer from interference from co-administered fluticasone propionate.