首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Stable mediated amperometric biosensors using a graphite electrode embedded with tetrathiafulvalene in silicone oil
Authors:Shishan Zhao  John HT Luong  
Institution:

Biotechnology Research Institute, National Research Council Canada, Montreal, Quebec, H4P 2R2, Canada Tel: (514) 496-6175; Fax: (514) 496-6265

Abstract:A new method has been developed to incorporate the mediator, tetrathiafulvalene (TTF), to the electrode/solution interface of an amperometric biosensor. TTF was dissolved in methylphenyl polysiloxane (silicone oil) and embedded in a graphite disc electrode. The mediator was able to diffuse to the electrode surface at an electrocatalytically significant speed. The storage of TTF in the inert polysiloxane provided a long-lasting and stable mediator supply.

TTF-silicone oil electrodes with immobilized glucose oxidase, xanthine oxidase, or amino acid oxidase exhibited sensitive, fast and reproducible responses. The glucose oxidase electrode was very stable for at least 2 months when stored at 4°C. Together with flow injection analysis (FIA), the enzyme electrodes were reused for at least 500 repeated analyses during a 25 h operation without losing their initial activity.

Keywords:mediated biosensor  tetrathiafulvalene  silicone oil  graphite electrode  glucose oxidase  xanthine oxidase  and L-amino acid oxidase