首页 | 本学科首页   官方微博 | 高级检索  
     


Flow cytometric fluorescence emission spectrum analysis of Hoechst-33342-stained DNA in chicken thymocytes
Authors:J V Watson  A Nakeff  S H Chambers  P J Smith
Abstract:Hoechst-33342-stained chicken thymocytes were analysed simultaneously on two fluorescence wavelength bands (green and violet) in our custom-built flow cytometer, and two major subsets were identified. In one subset (33% of the total) the emission spectrum remained constant with time, with little change in the respective green and violet fluorescence intensities. In the other subset (42% of the total) the green fluorescence increased during staining, resulting in a considerable change in the green-to-violet ratio, due to a change in the "shape" of the fluorescence emission with time. The data indicate that two binding sites, or two types of binding at the same site, exist in DNA for this dye and that these have different binding energies and, consequently, different fluorescence emission properties.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号