首页 | 本学科首页   官方微博 | 高级检索  
     


Solid phase extraction procedure for urinary organic acid analysis by gas chromatography mass spectrometry
Authors:Liu Aiping  Kushnir Mark M  Roberts William L  Pasquali Marzia
Affiliation:ARUP Institute for Clinical and Experimental Pathology, 500 Chipeta Way, Salt Lake City, UT 84108, USA. liua@aruplab.com
Abstract:We have developed a solid phase extraction procedure for the detection of organic acids by GC-MS using a strong anion exchange column (Sep-Pak Vac RC, Accell Plus QMA cartridge). Extraction efficiencies of 25 organic acids were established by analyzing standards in water based solutions. High extraction efficiencies (90 to 100%) were found for many of the compounds studied. We estimated the limit of detection for 48 organic acids and glycine conjugates. They were below 5 nmole with the exception of malonic and oxalic acids and mevalonic acid lactone. This method provides the advantage of higher recoveries for a wide range of compounds of interest and therefore can be a potential alternative to liquid-liquid extraction for organic acid screening. It is especially sensitive for the detection of some polar compounds, such as 3-OH-glutaric and N-acetylaspartic acids.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号