Extremely low frequency magnetic fields originating from equipment used for assisted reproduction, umbilical cord and peripheral blood stem cell transplantation, transfusion, and hemodialysis |
| |
Authors: | Moriyama Kosei Sato Hiroyuki Tanaka Kenji Nakashima Yoshiyuki Yoshitomi Kuniaki |
| |
Affiliation: | Department of Medicine, Fukuoka Dental College, 2-15-1 Tamura, Sawara-ku, Fukuoka, Japan. moriyama@college.fdcnet.ac.jp |
| |
Abstract: | The extremely low-frequency (ELF) magnetic fields (MFs) originating from equipment used for assisted reproduction, umbilical cord-blood and peripheral-blood stem cell transplantation, transfusion, and hemodialysis were measured. The ELF-MF values were 0.1-1.2 microT on clean benches, <0.1-8.0 microT on inverted microscopes, <0.1-13.6 mmicroT in CO2 incubators, 4.3-11.5 microT in centrifuges, 0.4-18.8 microT in programmed freezers, <0.1-0.3 microT in deep freezers, 0.3-3.1 microT on cell separators, and 0.2-0.9 microT in hemodialysers. Frequencies of MFs were nominally 60 Hz, but some devices showed non-sinusoidal 120 Hz. Such MFs can be reduced by shielding the sources or altering the protocols employed. |
| |
Keywords: | medical technology medical devices dosimetry Permalloy amorphous |
本文献已被 PubMed 等数据库收录! |
|