Disposable amperometric immunosensor system for rabbit IgG using a conducting polymer modified screen-printed electrode |
| |
Authors: | Darain Farzana Park Sang-Un Shim Yoon-Bo |
| |
Institution: | Department of Chemistry, Pusan National University, 609-735, Pusan, South Korea. |
| |
Abstract: | A disposable and mediatorless immunosensor based on a conducting polymer (5,2':5'2"-terthiophene-3'-carboxylic acid) coated screen-printed carbon electrode has been developed using a separation-free homogeneous technique for the detection of rabbit IgG as a model analyte. Horseradish peroxidase (HRP) and streptavidin were covalently bonded with the polymer on the electrode and biotinylated antibody was immobilized on the electrode surface using avidin-biotin coupling. This sensor was based on the competitive assay between free and labeled antigen for the available binding sites of antibody. Glucose oxidase was used as a label and in the presence of glucose, H(2)O(2) formed by the analyte-enzyme conjugate was reduced by the enzyme channeling via HRP bonded on the electrode. The catalytic current was monitored amperometrically at -0.35 V vs. Ag/AgCl and this method showed a linear range of RIgG concentrations from 0.5 to 2 microg/ml with standard deviation +/-0.0145 (n=4). Detection limit was determined to be 0.33 microg/ml. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|