DIRECT COUNTING AND SIZING OF MITOCHONDRIA IN SOLUTION
Authors:Adrian R L Gear and  Jana M Bednarek
Institution:From the Department of Biochemistry, The University of Virginia, Charlottesville, Virginia 22901
Abstract:Resistive particle counting has been developed for the accurate sizing and counting of mitochondria in solution. The normal detection limit with a 30 µ aperture is 0.48 µ diameter, or 0.056 µ3 particle volume The mean volume of rat liver mitochondria was 0.42 µ3 or 0.93 µ in diameter. The average value for numbers of particles per milligram of mitochondrial protein was 4.3 x 103, and per gram of rat liver was about 11 x 1010. These values compare satisfactorily with those derived by light microscopy and electron microscopy. The mean volume for mitochondria from rat heart was 0 60 µ3 and from rat kidney cortex, 0.23 µ3. These values agree within 15% of those determined by electron microscopy of whole tissue. Mitochondrial fragility and contaminating subcellular organelles were shown to have little influence on the experimentally determined size distributions The technique may be applied to rapid swelling studies, as well as to estimations of the number and size of mitochondria from animals under different conditions such as liver regeneration and hormonal, pathological, or drug-induced states Mitochondrial DNA, RNA, cytochrome c-oxidase, cytochrome (a ÷ a3), and iron were nearly constant per particle over large differences in particle size. Such data may be particularly valuable for biogenesis studies and support the hypothesis that the net amount per particle of certain mitochondrial constituents remains constant during mitochondrial growth and enlargement
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号