Plasma L-Ergothioneine Measurement by High-Performance Liquid Chromatography and Capillary Electrophoresis after a Pre-Column Derivatization with 5-Iodoacetamidofluorescein (5-IAF) and Fluorescence Detection
Authors:Salvatore Sotgia  Elisabetta Pisanu  Gianfranco Pintus  Gian Luca Erre  Gerard Aime Pinna  Luca Deiana  Ciriaco Carru  Angelo Zinellu
Affiliation:1. Department of Biomedical Sciences, University of Sassari, Sassari, Italy.; 2. Department of Clinical and Experimental Medicine, University of Sassari, Sassari, Italy.; 3. Department of Chemistry and Medicinal Chemistry, University of Sassari, Sassari, Italy.; University of Helsinki, Finland,
Abstract:Two sensitive and reproducible capillary electrophoresis and high-performance liquid chromatography-fluorescence procedures were established for quantitative determination of L-egothioneine in plasma. After derivatization of L-ergothioneine with 5-iodoacetamidofluorescein, the separation was carried out by HPLC on an ODS-2 C-18 sperisorb column by using a linear gradient elution and by HPCE on an uncoated fused silica capillary, 50 µm id, and 60 cm length. The methods were validated and found to be linear in the range of 0.3 to 10 µmol/l. The limit of quantification was 0.27 µmol/l for HPCE and 0.15 µmol/l for HPLC. The variations for intra- and inter-assay precision were around 6 RSD%, and the mean recovery accuracy close to 100% (96.11%).