Application of high-performance liquid chromatography with a chemiluminescence detection system to determine catecholamines in urine |
| |
Authors: | Shin-Ichiro Kobayashi Jun Sekino Kazumasa Honda Kazuhiro Imai |
| |
Institution: | Department of Analytical Chemistry, Faculty of Pharmaceutical Sciences, University of Tokyo, Hongo 7-3-1, Bunkyo-ku, Tokyo 113, Japan |
| |
Abstract: | Chemiluminescence generated with the reaction of bis(2,4,6-trichlorophenyl)oxalate and hydrogen peroxide was applied to a detection system for high-performance liquid chromatography to determine fluorescamine-labeled catecholamines. The sensitivity of the chemiluminescence detection system with 25 fmol of detection limit was approximately 20 times higher than that of a conventional fluorescence detection system. Norepinephrine and dopamine in human urine were determined by the use of the new high-performance liquid chromatography detection system with the coefficient of variation of less than 4.0%. Good correlations (r = 0.998 for norepinephrine and r = 0.999 for dopamine) were obtained between the values by the present method and the conventional method. |
| |