首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Rapid Prototyping of a High Sensitivity Graphene Based Glucose Sensor Strip
Authors:Farshad Tehrani  Lisa Reiner  Behzad Bavarian
Institution:1. Keck’s Advanced Materials Laboratory, Manufacturing Systems Engineering & Management, California State University of Northridge, Northridge, California, United States of America.; 2. Corrosion Research Laboratory, Manufacturing Systems Engineering & Management, California State University of Northridge, Northridge, California, United States of America.; US Naval Reseach Laboratory, UNITED STATES,
Abstract:A rapid prototyping of an inexpensive, disposable graphene and copper nanocomposite sensor strip using polymeric flexible substrate for highly sensitive and selective nonenzymatic glucose detection has been developed and tested for direct oxidization of glucose. The CuNPs were electrochemically deposited on to the graphene sheets to improve electron transfer rates and to enhance electrocatalytic activity toward glucose. The graphene based electrode with CuNPs demonstrated a high degree of sensitivity (1101.3±56 μA/mM.cm2), excellent selectivity (without an interference with Ascorbic Acid, Uric Acid, Dopamine, and Acetaminophen), good stability with a linear response to glucose ranging from 0.1 mM to 0.6 mM concentration, and detection limits of 0.025 mM to 0.9 mM. Characterization of the electrodes was performed by scanning electron microscopy (FESEM and SEM). The electrochemical properties of the modified graphene electrodes were inspected by cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), and amperometry.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏