Infrared absorbance spectroscopy of aqueous proteins: Comparison of transmission and ATR data collection and analysis for secondary structure fitting
Authors:Marco Pinto Corujo  Meropi Sklepari  Dale L. Ang  Mark Millichip  Andrew Reason  Sophia C. Goodchild  Paul Wormell  Don Praveen Amarasinghe  Viv Lindo  Nikola P. Chmel  Alison Rodger
Affiliation:1. Department of Chemistry, University of Warwick, Coventry, UK;2. MOAC and MAS Centres for Doctoral Training, University of Warwick, Coventry, UK;3. School of Science and Health, Western Sydney University, Penrith, New South Wales, Australia;4. BioPharmaSpec Ltd., Jersey, UK;5. Department of Molecular Sciences, Macquarie University, North Ryde, New South Wales, Australia;6. MedImmune Ltd, Cambridge, UK
Abstract:Attenuated total reflectance (ATR) infrared absorbance spectroscopy of proteins in aqueous solution is much easier to perform than transmission spectroscopy, where short path‐length cells need to be assembled reproducibly. However, the shape of the resulting ATR infrared spectrum varies with the refractive index of the sample and the instrument configuration. Refractive index in turn depends on the absorbance of the sample. In this work, it is shown that a room temperature triglycine sulfate detector and a ZnSe ATR unit can be used to collect reproducible spectra of proteins. A simple method for transforming the protein ATR spectrum into the shape of the transmission spectrum is also given, which proceeds by approximating a Kramers‐Krönig–determined refractive index of water as a sum of four linear components across the amide I and II regions. The light intensity at the crystal surface (with 45° incidence) and its rate of decay away from the surface is determined as a function of the wave number–dependent refractive index as well as the decay of the evanescent wave from the surface. The result is a single correction factor at each wave number. The spectra were normalized to a maximum of 1 between 1600 cm?1 and 1700 cm?1 and a self‐organizing map secondary structure fitting algorithm, SOMSpec, applied using the BioTools reference set. The resulting secondary structure estimates are encouraging for the future of ATR spectroscopy for biopharmaceutical characterization and quality control applications.
Keywords:attenuated total reflectance  ATR correction  refractive index  depth of penetration
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号