Comparative study of chemical pathology sample collection tubes at the largest hospital in South Africa
Authors:Siyabonga P. Khoza  Sarah Ford  Ernest P. Buthelezi  Donald M. Tanyanyiwa
Affiliation:1. University of the Witwatersrand, Johannesburg, South Africa ; 2. National Health Laboratory Service, Department of Chemical Pathology, Johannesburg, South Africa ; 3. Gauteng Department of Health, Johannesburg, South Africa ; 4. Sefako Makgatho University of Health Sciences, Pretoria, South Africa
Abstract:BackgroundBarricorTM Lithium heparin plasma tubes are new blood tubes that have been introduced to overcome the effects of gel in serum separator tubes (SST) and the shortcomings of standard Lithium heparin plasma. We aimed to evaluate BarricorTM tubes as an alternative to serum separator tubes and compare the stability between the tubes.MethodsForty-four paired samples were collected using both BarricorTM and SST. We compared five analytes at baseline (<6 h) and after every 24 h using the PassingBablok and Bland-Altman plots. Aspartate aminotransferase (AST), potassium (K), phosphate (PO4) , lactate dehydrogenase (LDH), and creatinine were analysed in both tubes. We calculated the percentage difference for each analyte between the baseline and time intervals to assess analyte stability. The percentage difference was compared to the desirable specification for bias and reference change value (RCV).ResultsAll analytes were comparable at baseline. Statistical differences (p<0.001) became evident after 24 h. PO4, K, and creatinine had a mean difference that exceeded the desirable specification for bias (-9.59%, - 9.35%, and -4.59%, respectively). Potassium was stable up to 24 h in both tubes. LDH showed better stability in SST (144 h vs 96 h). PO4 concentrations were more stable in both tubes with the SST (96 h vs 72 h). Creatinine and AST had the longest stability in both tubes compared to other analytes (144 h).ConclusionsData demonstrated variability and similarities in analyte concentrations and stability, respectively, in both tubes.
Keywords:BarricorTM   blood collection devices   serum separator tubes
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号