Abstract: | A high-performance liquid chromatographic procedure is reported for reproducibly and sensitively quantitating caffeine and its N-demethylated metabolite paraxanthine in micro-samples. A 5-μm reversed-phase radial compression column and 214-nm fixed wavelength ultraviolet detector were used to attain a sensitivity sufficient to quantitate these compounds at concentratios as low as 80 ng/ml using only 25 μl of sample. The assay is applicable to microliter samples of whole blood, serum, plasma, saliva, amniotic, cerebro-spinal and gastric fluids such as might be obtained in studies involving small animals or neonates. The utility of the assay is illustrated with caffeine and paraxanthine levels measured in several maternal and fetal fluids following constant-rate intravenous infusion of caffeine into a rabbit throughout pregnancy. |