Fluorometric determination of adenine nucleotides and adenosine by ion-paired, reverse-phase, high-performance liquid chromatography |
| |
Authors: | A Ramos-Salazar A D Baines |
| |
Abstract: | A sensitive and specific assay for measurement of adenine nucleotides and adenosine by paired-ion high-performance liquid chromatography is described. The 1,N6-ethenoderivatives of ATP (epsilon-ATP), ADP (epsilon-ADP), AMP (epsilon-AMP), and adenosine (epsilon-Ado), formed by reaction with chloroacetaldehyde at 37 degrees C, were separated under isocratic conditions in 20 min. These compounds are strongly fluorescent at an emission wavelength of 280 nm, rendering a lowest detection limit of 2-5 pmol per injection. The detector responded linearly over the measured ranges (5-100 pmol for epsilon-Ado and 5-4000 pmol for nucleotides). Specificity was confirmed enzymatically. alpha, beta-Methyleneadenosine 5'-diphosphate could be used as an internal standard for measurement of the nucleotides. Significant amounts of NADH appeared as a separate peak in hypoxic tissue. Recoveries from snap-frozen kidney were 88, 92, 76, and 63% for AMP, ADP, ATP, and adenosine, with SD for recovery of 1.0, 10.5, 8.3, and 5.6%, respectively. This method was successfully used to measure adenine nucleotides and adenosine in oxygenated and hypoxic perfused rat kidneys. |
| |