Determination of ragaglitazar,a novel dual acting peroxisome proliferator-activated receptor (PPAR) alpha and -gamma agonist,in human plasma by high-performance liquid chromatography coupled with tandem mass spectrometry |
| |
Authors: | Andersen Michael P Nielsen Karin K |
| |
Institution: | Department of Pharmacokinetics, Research and Development, Novo Nordisk A/S, Novo Nordisk Park, DK-2760 Maaloev, Denmark. mipa@novonordisk.com |
| |
Abstract: | A sensitive and specific LC/MS/MS method has been developed and validated for determination of ragaglitazar (NNC 61-0029 or DRF 2725) in human plasma. After solid-phase extraction (SPEC((R)) PLUS C(8)) of plasma, separation was performed on a Symmetry Shield RP8 column (mobile phase: acetonitrile: 10 mM ammonium acetate, pH 5.6 (40:60 v/v)). Two ranges were validated having LLOQs of either 0.500 or 100 ng/ml and linearity up to either 500 or 50000 ng/ml. The intra-assay precision and accuracy were 1.1% to 15.7% and 85.8% to 118.2% (range 0.500-500 ng/ml) and 2.0% to 8.8% and 92.9% to 104.8% (range 100-50000 ng/ml). The method was applied for determination of ragaglitazar in plasma from phase 1 and 2 clinical studies. |
| |
Keywords: | Ragaglitazar |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|