首页 | 本学科首页   官方微博 | 高级检索  
     


Use of a copper-phthalocyanine membrane electrode for rapid preliminary detection of polycyclic mutagens
Authors:R Tomoda  S Kusunoki  K Nakashima  T Matsunaga
Abstract:Preliminary screening of polycyclic mutagens is achieved within 20 min by using a biomimetic electrode composed of an oxygen electrode and a copper-phthalocyanine membrane. When benzo[alpha]pyrene (0.05 mM) was added to the buffer solution in the presence of 0.98 M hydrogen peroxide, the current of the phthalocyanine electrode decreased. A linear relationship was obtained between the current decrease and the benzo[alpha]pyrene concentration over the range 0.19-0.60 mM. The minimum measurable concentration for benzo[alpha]pyrene was 0.01 mM. Such responses were not obtained for other organic compounds such as alcohol, ether, n-hexane and cyclohexane. The copper-phthalocyanine membrane electrode has selectively detected polycyclic mutagens such as amino acid pyrolysis products. The current decrease was 1.18-1.46 microA when 0.05 mM amino acid pyrolysis products were employed.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号