首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Stabilization and determination of a PPAR agonist in human urine using automated 96-well liquid-liquid extraction and liquid chromatography/tandem mass spectrometry
Authors:Groff Michelle  Riffel Kerry  Song Hengchang  Lo Man-Wai
Institution:Department of Drug Metabolism, Merck Research Laboratories, WP75A-303, West Point, PA 19486, USA. michelle_groff@merck.com
Abstract:Two stability challenges were encountered during development of an urine assay for a proliferator-activated receptor (PPAR) agonist, I (2-{5,7-dipropyl-3-(trifluoromethyl)-1,2-benzisoxazol-6-yl]oxy}-2-methyl propionic acid), indicated for the treatment of Type II diabetes. First, the analyte was lost in urine samples due to adsorption on container surface which is a common problem during clinical sample handling. Secondly, the acylglucuronide metabolite (III), a major metabolite of I, displayed limited stability and effected the quantitation of parent drug due to the release of I through hydrolysis. Therefore, a clinical collection procedure was carefully established to stabilize I and its acylglucuronide metabolite, III, in human urine. The metabolite was not quantitated with this method. The urine samples are treated with bovine serum albumin (BSA) equal to 1.75% of the urine volume and formic acid equal to 1% of urine volume. Compound (I) and internal standard (II) were extracted from urine with 1 mL ethyl acetate using a fully automated liquid-liquid extraction in 96-well plate format. The analytes are separated by reverse phase high-performance liquid chromatography (HPLC) with tandem mass spectrometry in multiple-reaction-monitoring (MRM) mode used for detection. The urine method has a lower limit of quantitation (LLOQ) of 0.05 ng/mL with a linearity range of 0.05-20 ng/mL using 0.05 mL of urine. The method was validated and used to assay urine clinical samples.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号