Rapid high-performance liquid chromatographic method for the simultaneous determination of retinol, α-tocopherol and β-carotene in human plasma and low-density lipoproteins
Authors:E Gimeno  A I Castellote  R M Lamuela-Ravents  M C de la Torre-Boronat  M C Lpez-Sabater
Institution:E. Gimeno, A. I. Castellote, R. M. Lamuela-Raventós, M. C. de la Torre-Boronat,M. C. López-Sabater,
Abstract:A reversed-phase HPLC method with diode-array detection was used to simultaneously determine retinol, α-tocopherol and β-carotene in human plasma and low-density lipoproteins. An aliquot of sample was de-proteinized with ethanol containing α-tocopherol acetate as internal standard, and the analytes were extracted twice with hexane. The solvent was evaporated to dryness under a stream of nitrogen and the residue was redissolved in methanol to be injected directly into the HPLC system. A multiple solvent system based on methanol, butanol and water at a flow-rate of 2 ml/min and held at 45°C provided clear separation of these compounds in only 8 min. The method showed good linearity, precision and accuracy for all compounds. Owing to its simplicity, this method may be useful in routine clinical and epidemiological work.
Keywords:Vitamins  Retinol  α  -Tocopherol  β  -Carotene