The Effects of Computed Tomography with Iterative Reconstruction on Solid Pulmonary Nodule Volume Quantification
Authors:Martin J. Willemink  Jaap Borstlap  Richard A. P. Takx  Arnold M. R. Schilham  Tim Leiner  Ricardo P. J. Budde  Pim A. de Jong
Affiliation:1. Utrecht University Medical Center, Department of Radiology, Utrecht, The Netherlands.; 2. Gelre Hospital, Department of Radiology, Apeldoorn, The Netherlands.; The University of Chicago, United States of America,
Abstract:

Background

The objectives of this study were to evaluate the influence of iterative reconstruction (IR) on pulmonary nodule volumetry with chest computed tomography (CT).

Methods

Twenty patients (12 women and 8 men, mean age 61.9, range 32–87) underwent evaluation of pulmonary nodules with a 64-slice CT-scanner. Data were reconstructed using filtered back projection (FBP) and IR (Philips Healthcare, iDose4-levels 2, 4 and 6) at similar radiation dose. Volumetric nodule measurements were performed with semi-automatic software on thin slice reconstructions. Only solid pulmonary nodules were measured, no additional selection criteria were used for the nature of nodules. For intra-observer and inter-observer variability, measurements were performed once by one observer and twice by another observer. Algorithms were compared using the concordance correlation-coefficient (pc) and Friedman-test, and post-hoc analysis with the Wilcoxon-signed ranks-test with Bonferroni-correction (significance-level p<0.017).

Results

Seventy-eight nodules were present including 56 small nodules (volume<200 mm3, diameter<8 mm) and 22 large nodules (volume≥200 mm3, diameter≥8 mm). No significant differences in measured pulmonary nodule volumes between FBP, iDose4-levels 2, 4 and 6 were found in both small nodules and large nodules. FBP and iDose4-levels 2, 4 and 6 were correlated with pc-values of 0.98 or higher for both small and large nodules. Pc-values of intra-observer and inter-observer variability were 0.98 or higher.

Conclusions

Measurements of solid pulmonary nodule volume measured with standard-FBP were comparable with IR, regardless of the IR-level and no significant differences between measured volumes of both small and large solid nodules were found.