Determination of rifabutin by high-performance liquid chromatography using on-line concentration and column switching
Authors:Helga Bartels  Rainer Bartels
Institution:Borstel Research Center, Parkallee 4a, D-23845 Borstel, Germany
Abstract:A simple HPLC method has been developed that allows the sensitive determination of rifabutin (RBT) in human serum using on-line concentration and column switching. After pretreatment of the serum with acetonitrile and centrifugation, the samples were applied to a concentration column (CC) (Zorbax CN). Washing with phosphate buffer-methanol removed most of the contaminating substances. Via a six-port valve the CC was switched to the analytical mode. RBT was separated on a Chromspher RP 8 column (acetonitrile-phosphate buffer pH 7.4/sodium chloride) and determined photometrically at 278 nm. The lower limit of quantification for 200 μl serum precipitated with 200 μl acetonitrile and after injection of 2×150 μl was 33 μg/l and linearity was observed up to 27 mg/l. Different modes of sample application (single, repeated, and different injection volume portions), as well as washing time, cycle time and different CC materials were investigated.
Keywords:Rifabutin
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号