Abstract:Sumatriptan succinate (the analyte) and naloxone (the internal standard) were extracted from plasma with a solid-phase extraction technique. Chromatography and detection were performed by isocratic reversed-phase high-performance liquid chromatography with coulometric end-point detection. The standard curve was linear over the range 0–100 ng/ml of sumatriptan succinate in plasma. The reproducibility (as defined by the coefficient of variation, C.V.) over the range of the standard curve was 4.9–7.3%. The recovery averaged 83%. The sensitivity was 0.25 ng of sumatriptan on column (allowing a concentration of 0.5 ng/ml to be determined from a 1-ml plasma sample volume). Plasma profiles of the analyte following subcutaneous (s.c.) administration in eight normal male volunteers, are presented.