首页 | 本学科首页   官方微博 | 高级检索  
     


Mechanical aspects of the development of artificial heart valves
Authors:A Bolz  M Schaldach  W Winter
Affiliation:Zentralinstitut für Biomedizinische Technik, Universit?t Erlangen-Nürnberg.
Abstract:The development of an antithrombogenic coating permits a hybrid design for artificial heart valves. A substrate material optimized for its application is coated to meet the electrochemical requirements of improved hemocompatibility. But future progress in artificial heart valves requires an improvement in design as well as of the material. The basis of both aspects is the determination of such fundamental mechanical properties as the elasticity and plasticity of the valve ring and the deformation and fraction behaviour of the occluder. Analytical and numerical calculations of various different models result in different requirements for the substrate of ring and occluder. A combination of high elastic temper and low resistance to flow requires a ring material with a Young's modulus of 40 GPa or more, and a 0.2% proof stress to (Young's modulus)2/3 ratio of 0.3 MPa1/3. The best occluder materials should have a Young's modulus of more than 50 GPa and a flexural strength of at least 800 MPa. On the basis of these criteria, a heart valve consisting of a TiA15Fe2,5 ring and occluders made partially stabilized zirconia is introduced.