首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Determination of acyclovir by ultrafiltration and high-performance liquid chromatography
Authors:Peter Nebinger  Marlies Koel
Abstract:A simple, sensitive and rapid high-performance liquid chromatographic (HPLC) procedure to determine total serum acyclovir concentrations is described. The assay involves a heat inactivation step at 56°C to prevent risk of infection, ultrafiltration as a pretreatment step prior to ion-pair reversed-phase liquid chromatography using guanosine as internal standard, and ultraviolet detection at 254 nm. This method has excellent recovery (97–100%), linearity (0.5–100 mg/l) and precision (1.2–8.0% coefficient of variation). The detection limit is 50 μg/l. The assay proved to be suitable for therapeutic drug monitoring of acyclovir.
Keywords:
本文献已被 ScienceDirect 等数据库收录!