Validation of the measurement of low concentrations of 5-hydroxytryptamine in plasma using high performance liquid chromatography |
| |
Authors: | Atkinson Wendy Lockhart Stephen J Houghton Lesley A Keevil Brian G |
| |
Affiliation: | Department of Medicine, South Manchester University Hospitals NHS Trust, UK. |
| |
Abstract: | A sensitive and rapid assay is described for the measurement of low concentrations of 5-hydroxytryptamine (5-HT) present in human platelet-depleted plasma (PDP) using reverse-phase high performance liquid chromatography (HPLC) with fluorimetric detection. With an analysis time of 12 min, this method is particularly useful for large-scale clinical trials investigating small differences in PDP 5-HT concentrations in conditions such as functional gastrointestinal disorders (FGID). The limit of detection and quantification were 1 and 3 nmol/l, respectively, and the calibration curve linear between 1 and 1000 nmol/l. The within-day and between-day precision were 4.3 and <13.6%, respectively. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|