RNAA as compared to ICP-MS for the analysis of normal human serum
Authors:Dams  R
Institution:1.Lab. Analytical Chemistry, Institute for Nuclear Sciences, University Gent, Proeftuinstraat 86, B-9000, Gent, Belgium
;
Abstract:The merits of radiochemical neutron activation analysis (RNAA) and inductively coupled plasma mass spectrometry (ICP-MS) are critically discussed for the determination of trace and ultratrace elements in normal human serum. For RNAA, two semiautomated procedures, allowing the determination of up to 18 elements, are briefly described. ICP-MS has a series of interesting features for the determination of trace elements. Matrix and spectral interferences can mostly be avoided or corrected for. After a simple 5- or 10-fold dilution and addition of an internal standard, more than 20 elements can be measured precisely and accurately.
Keywords:Index Entries" target="_blank">Index Entries  Human serum  radiochemical neutron activation analysis  inductively coupled plasma mass spectrometry