首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Airflows In and Around Linear Downflow 'Safety' Cabinets
Authors:R P CLARK  B J MULLAN
Institution:Laboratory for Aerobiology, Division of Hospital Infection, Clinical Research Centre, Harrow HA1 3UJ, Middlesex, England
Abstract:Schlieren photography has been used to visualize the airflows in and around three re-circulating linear downflow cabinets. The results show that there is an air 'bulge' at the front opening that allows air from within the cabinet to spill out towards the operator. Hand movements within, and activity just outside, cause air to be expelled from these cabinets. Bunsen burners within the cabinets destroy the linear nature of the downflowing air streams. The results indicate that the operator protection afforded by these cabinets may be much less than is generally recognized.