HPLC-analysis of algal pigments: comparison of columns,column properties and eluents
Authors:Hans Schmid  Hans Bernd Stich
Institution:(1) Institut für Seenforschung, Untere Seestr. 81, 88085, Langenargen, Germany
Abstract:The effects of columns (Nucleosil C18ODS, MZ-PAH, YMC-PACK C30), column properties (inner diameters of 4 mm, 3 mm and 2 mm, pore-width 10 nm and 30 nm) and eluents (methanol, acetonitrile, acetone, water) were tested on the separation of algal pigments. The length of columns was 250 mm and particle size was 5 μm. Flow rates and gradients were adjusted to optimize peak separation; remaining chromatographic conditions were kept constant. The resolution of chromatographic systems was tested with pigment standards and various algal cultures. Total flow rate and retention times decreased with decreasing inner diameter, whereas pressure, sensitivity and peak-width increased. Pore width had negligible effects on the chromatographic separation of pigments under the test conditions. Only with acetonitrile as eluent were all the taxonomically important pigments resolved adequately: zeaxanthin (Cyanophyceae), lutein (Chlorophyceae), fucoxanthin (Bacillariopyceae), alloxanthin (Cryptophyceae), peridinin (Dinophyceae).
Keywords:HPLC  method  algal pigments  chlorophyll  carotenoid