首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Measurement of the size distribution of human erythrocytes by a sedimentation method
Authors:A C Groom  J C Anderson
Abstract:The distribution of sedimentation velocities was determined, by a photoelectric method, for human erythrocytes at low concentrations in Ringer solution. The light absorption at 414 nm was measured, as a function of time, 10 mm below the top of the column. From the frequency distribution of cell velocities that of Rs √ρ-σ was found; Rs being the Stokes' radius, ρ the cell density and σ the density of the solution. Cell density was measured by the phthalate method and the mean Stokes' radius was found to be 2.58 μm. The size distributions showed some skewness but were in good general agreement with those measured by Celloscope counter, and with reported measurements from photomicrographs of cells in hanging drop suspensions. The skewness was much less than that encountered with electric sensing zone instruments (e.g. Celloscope) and the sedimentation method, being based on entirely different premises, provides an important check on such data. The skewness is due to a bias in the orientation of human erythrocytes during sedimentation. This bias may be a characteristic of biconcave cells; it could be absent in many species and reliable measurements of size distribution would then be obtained.