High-performance liquid chromatographic assay for cefepime in serum |
| |
Authors: | H. Elkhaï li, L. Linger, H. Monteil,F. Jehl |
| |
Affiliation: | Laboratory of Pharmacokinetics, Institute of Bacteriology, University Hospital Center, 3 Rue Koeberlé, 67000, Strasbourg, France |
| |
Abstract: | A simple, rapid, specific and sensitive high-performance liquid chromatographic method was developed for the determination of cefepime 1-[[6R, 7R)-7-[2-(2-amino-4-thiazolyl)glyoxylamido]-2-carboxy-8-oxo-5-thia-1-azabicyclo-[4.2.0] oct-2-en-3-yl]methyl]-1-methylpyrrolidinium hydroxide, inner salt, 72-(Z)-(O-methyloxime) in human serum. Separation was achieved on a reversed-phase Ultrasphere XL-ODS column (75×4.6 mm I.D.). The mobile phase was 7% acetonitrile in 20 mM ammonium acetate (pH 4). Cefepime eluted in the range of 1.8–2.2 min. Detection was by UV absorbance at 254 nm. The lower limit of quantitation of cefepime in plasma was 0.5 μg/ml. The average absolute recovery was 106.2±2.1%. The linear range was from 0.1 to 50 μg/ml, with a correlation coefficient greater than 0.999. The within-day C.V.s for human samples were 4.9 and 2.3% for 1 and 50 μg/ml, respectively. The between-day C.V.s for human serum samples were 14.5, 7.4 and 6.7 for 1, 25 and 50 μg/ml, respectively. Cefepime was found to be unstable in serum at room temperature. For delayed assay, samples must be stored at −80°C. |
| |
Keywords: | Cefepime |
本文献已被 ScienceDirect 等数据库收录! |
|