首页 | 本学科首页   官方微博 | 高级检索  
     


Design and construction of a preparative-scale dynamic field gradient focusing apparatus
Authors:Tracy Noah I  Huang Zheng  Ivory Cornelius F
Affiliation:School of Chemical Engineering and Bioengineering, Washington State University, Pullman, WA 99164, USA.
Abstract:A linear model is used to show that dynamic field gradient focusing (DFGF) can be scaled to preparative capacity, approximately O (10 mgs). This paper explains how the preparative-scale DFGF apparatus was designed and fabricated. Scaled-down experiments and mathematical modeling guided material selection and design changes during construction to increase the probability that the prototype preparative-scale DFGF apparatus would perform as intended. The finished prototype successfully focused bovine hemoglobin from an initial concentration of 6.82 to 15 mg/mL and allowed for 86% recovery of injected protein.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号