首页 | 本学科首页   官方微博 | 高级检索  
   检索      


Rapid tilt-series acquisition for electron cryotomography
Authors:Georges Chreifi  Songye Chen  Lauren Ann Metskas  Mohammed Kaplan  Grant J Jensen
Institution:1. Division of Biology and Biological Engineering, California Institute of Technology, Pasadena, CA 91125, United States;2. Beckman Institute, California Institute of Technology, Pasadena, CA 91125, United States;3. Howard Hughes Medical Institute, California Institute of Technology, Pasadena, CA 91125, United States
Abstract:Using a new Titan Krios stage equipped with a single-axis holder, we developed two methods to accelerate the collection of tilt-series. We demonstrate a continuous-tilting method that can record a tilt-series in seconds, but with loss of details finer than ~4?nm. We also demonstrate a fast-incremental method that can record a tilt-series several-fold faster than current methods and with similar resolution. We characterize the utility of both methods in real biological electron cryotomography workflows. We identify opportunities for further improvements in hardware and software and speculate on the impact such advances could have on structural biology.
Keywords:Cryo-electron tomography  Electron cryotomography  Cryo-EM  Tilt-series