Comparison of three different methods for measurement of tissue platinum level
Authors:Takeshi Minami  Masayo Ichii  Yuiko Okazaki
Affiliation:1. Faculty of Pharmaceutical Sciences, Kinki University, 3-4-1 Kowakae, Higashi-Osaka, 577, Osaka, Japan
Abstract:We attempted to make a comparison of three methods for tissue platinum; atomic absorption spectrometry (AAS), inductively coupled plasma atomic emission spectrometry (ICP-AES), and inductively coupled plasma mass spectrometry (ICP-MS). The determination lim its were 0.05 ng/mL on ICP-MS, 50 ng/mL on ICP-AES, and 200 ngJmL on AAS, and the recovery rates were 97.7 ± 6.9% on ICP-MS, 69.0 ± 3.0% on ICP-AES, and 102.4 ± 4.0% on AAS, respectively. Plat inum was detected by ICP-AES and ICP-MS in human vertebrae, but the level was higher by ICP-AES than by ICP-MS. In the mouse kid ney treated with cisplatin, platinum was detected by ICP-MS, but not by ICP-AES. As cadmium gives the absorption peak close to plat inum, cadmium was measured together with platinum by ICP-AES in the vertebrae. From these, ICP-MS is the most sensitive for measure ment at tissue platinum. The sensitivity of ICP-AES looks worse for measuring the tissue platinum, and it is necessary to take care of the contaminant of metals, especially cadmium. AAS is not suitable for measurement of tissue platinum as in the vertebrae and kidneys, because platinum was not detectable by AAS.
Keywords: