Effect of flow baffles on the dialysate flow distribution of hollow-fiber hemodialyzers: a nonintrusive experimental study using MRI
Authors:Poh Churn K  Hardy Peter A  Liao Zhijie  Huang Zhongping  Clark William R  Gao Dayong
Affiliation:Department of Mechanical Engineering, University of Kentucky, Lexington, KY, USA.
Abstract:We used an innovative, nonintrusive MRI technique called the two-dimensional (2D) Phase-Contrast (2DPC) velocity-imaging technique to investigate the effect of flow baffles on the dialysate-side flow distribution in two different hollow-fiber hemodialyzers (A and B); each with flow rates between 200 and 1000 mL/min (3.33 x 10(-6) and 1.67 x 10(-5) m3/s). Our experimental results show that (1) the dialysate-side flow distribution was nonuniform with channeling flow occurred at the peripheral cross section of these hollow-fiber hemodialyzers, and (2) the existing designs of flow baffles failed to promote uniform dialysate-side flow distribution for all flow rates studies.