Interstitial fluid glucose time-lag correction for real-time continuous glucose monitoring |
| |
Authors: | D. Barry Keenan John J. Mastrototaro Stuart A. Weinzimer Garry M. Steil |
| |
Affiliation: | 1. Medtronic MiniMed, 18000 Devonshire Street, Northridge, CA 91325, USA;2. Department of Pediatrics, Yale School of Medicine, New Haven, CT 06520, USA;3. Children''s Hospital Boston, 300 Longwood Avenue, Boston, MA 02115, USA |
| |
Abstract: | Time lag between subcutaneous interstitial fluid and plasma glucose decreases the accuracy of real-time continuous glucose monitors. However, inverse filters can be designed to correct time lag and attenuate noise enabling the blood–glucose profile to be reconstructed in real time from continuous measurements of the interstitial-fluid glucose. We designed and tested a Wiener filter using a set of 20 sensor-glucose tracings (~30 h each) with a 1-min sample interval. Delays of 10 ± 2 min (mean ± SD) were introduced into each signal with additive Gaussian white noise (SNR = 40 dB). Performance of the filter was compared to conventional causal and non-causal seventh-order finite-impulse response (FIR) filters. Time lags introduced an error of 5.3 ± 2.7%. The error increased in the presence of noise (to 5.7 ± 2.6%) and attempts to remove the noise with conventional low-pass filtering increased the error still further (to 7.0 ± 3.5%). In contrast, the Wiener filter decreased the error attributed to time delay by ~50% in the presence of noise (from 5.7% to 2.60 ± 1.26%) and by ~75% in the absence of noise (5.3% to 1.3 ± 1%). Introducing time-lag correction without increasing sensitivity to noise can increase CGM accuracy. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|