首页 | 本学科首页   官方微博 | 高级检索  
     


Glucose sensor using immobilized whole cells of Pseudomonas fluorescens
Authors:Isao Karube  Satoshi Mitsuda  Shuichi Suzuki
Affiliation:(1) Research Laboratory of Resources Utilization, Tokyo Institute of Technology, Nagatsuta-cho Midori-ku, 227 Yokohama, Japan
Abstract:Summary Whole cells of Pseudomonas fluorescens which utilized mainly glucose were immobilized in collagen membrane. The microbial electrode consisted of a bacteria-collagen membrane and an oxygen electrode was developed for the determination of glucose. When the electrode was inserted in a sample solution containing glucose, the current of the electrode decreased markedly with time until a steady state was reached. The response time of the electrode was 10 min by the steady state method. A linear relationship was observed between the steady state current and the concentration of glucose below 20 mg l–1. The minimum concentration for determination was 2 mg of glucose per liter. The reproducibility of the current was examined using the same sample solution. The current was reproducible within ±6% of the relative error when a sample solution containing 10 mg {ie343-1} of glucose was employed. The standard deviation was 0.6 mg {ie343-2} in 20 experiments. The reusability of the glucose sensor was examined using the same sample solution (10 mg {ie343-3}). No decrease in current output was observed over a two week period and 150 assays. Glucose in molasses was determined with an average relative error of 10% by the microbial electrode sensor.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号